Question
Requirements and referencesโ
- GSPR 23
Documents reviewed by BSIโ
ifu.pdf
Question 1: GSPR 23.1(a) compliance and corrective action 4โ
Compliance to GSPR 23.1(a) and effective implementation of corrective action 4 are not fully demonstrated because:
a. The sample label and instructions in installation instructions in the IFU, ยง User Interface do not meet the requirements of GSPR 23.2 (e.g., do not include the name of the device, UDI, date of manufacture, intended purpose, indication the device is a medical device, etc.)
b. The instructions do not require the installer to provide users a link to the eIFU website (e.g., via the user interface, training, email, etc.).
Please provide the location of this information or provide updated documentation if necessary.
BSI observation
It is also unclear why the installation instructions reference GSPR 17.1.