Question
Requirements and references
- Article 86
- Article 61(11)
MDF reference
CER updates
Question
Requirements
- Article 86 states that the PSUR for Class IIa and IIb devices shall be updated at least every 2 years
- Article 61 (11) states that the clinical evaluation shall be updated based on clinical data obtained from the implementation of the post market plans
Observations
§18 of the CER does not seem to specify a regular cadence for the update frequency of the CER (e.g. every 2 years, etc)
Request
Please provide a clear update frequency cadence for the clinical evaluation as aligned to requirements in Articles 61 (11) and Article 86.