Question

Requirements and references​

  • Chapter VII Articles 83–89
  • Annex III 1(a) and (b)

MDF reference​

MDF4550 §3.3.1

Question​

Requirements​

  • Chapter VII Articles 83-89 describe the requirements for post market surveillance and vigilance
  • Annex III 1 (a) and (b)

Observation​

It remains unclear how all of the requirements of the referenced Annexes and Articles have been met.

For example (not exhaustive):

  • SOPs with sufficient detail to demonstrate effective and appropriate methods and processes to assess the collected data are not found
  • Specific and suitable indicators and threshold values to be used in the continuous reassessment of B/R and risk management are not found
  • Detail on effective and appropriate methods and tools to investigate complaints and analyze market-related experience from the field per Article 87 (sub-clauses 1, 2, 7, 8, 9 and 11) is not found
  • Details of specific methods of trending to manage the trend report per Article 88 is not found
  • Methods and protocols to communicate effectively with all relevant bodies is not found
  • Procedures detailing how the requirements of Articles 83 and 86 are not found (e.g. SOPs describing how requirements will be met)
  • Systematic procedures to identify and initiate appropriate measures and trace and identify devices for which corrective actions are needed do not seem to be detailed.

Request​

Please address all points above. Please provide traceability to how the plan meets the requirements for each point of Annex III 1 (b) and referenced requirements (such as Annex XIV 1 (a) and Article 83, 86, 87, 88). Please provide all relevant WIs/SOPs and page/section references to where the requirements have been met.