Skip to main content
QMSQMS
QMS
  • Welcome to your QMS
  • Quality Manual
  • Procedures
  • Records
  • Legit.Health Plus Version 1.1.0.0
  • Legit.Health Plus Version 1.1.0.1
  • Legit.Health Utilities
  • Licenses and accreditations
  • Applicable Standards and Regulations
  • BSI Non-Conformities
    • Technical Review
      • Round 1
        • M1: Diagnostic Function
          • Q1: IFU Performance Claims
          • Q2: Test Environment
          • Q3: Biofilm/Slough Verification
          • Q4: T377 Test Results
        • M2: Software V&V
        • N1: Information Supplied
        • N2: Usability
        • N3: Risk Management
    • BSI Non-Conformities
  • Pricing
  • Public tenders
  • BSI Non-Conformities
  • Technical Review
  • Round 1
  • M1: Diagnostic Function

M1: Diagnostic Function

Major Non-Conformity

Code: 2650005-202501-M1

Status: Open: awaiting manufacturer response

Requirements and references​

  • GSPR 1, 15.1, 17.2
  • Annex II, 6.1(a), (b); 6.2(f)
  • EN 62304
  • EN 82304-1

Documents reviewed by BSI​

  • ifu.pdf
  • master.csv
  • C515_T355/case-001/response.json
  • C537_T377/case-001/response.json

Regulatory context​

Internal working document

This section is for internal planning only. It will not be included in the final response to BSI.

BSI raised four questions under Major NC 2650005-202501-M1. The regulatory requirements cited are GSPR 1, 15.1, 17.2; Annex II 6.1(a), (b), 6.2(f); EN 62304; and EN 82304-1.

Cited requirements — detailed breakdown​

Each response must explicitly tie back to the specific requirement it satisfies. The cited requirements are:

RequirementWhat it demandsRelevant questions
GSPR 1Devices shall achieve intended performance, be safe and effective, with acceptable benefit-risk ratioQ1 (performance claims must be understandable)
GSPR 15.1Devices providing diagnostic information must indicate accuracy limits, precision, and stability; limits of accuracy shall be indicated by the manufacturerQ1 (IFU must clearly communicate accuracy metrics and their applicability)
GSPR 17.2Software shall be developed per state of the art: lifecycle, risk management, V&VQ2, Q3, Q4 (V&V evidence must be complete and correct)
GSPR 23.4(c)IFU must state residual risks and undesirable side-effectsQ1 (performance limitations must be communicated)
GSPR 23.4(e)IFU must state warnings and precautionsQ1 (accuracy limitations per condition/context)
GSPR 23.4(h)IFU must provide specifications for appropriate use, including degree of accuracy claimedQ1 (core of BSI's concern — accuracy claims need context)
MDR Annex II 6.1(a)Technical documentation must include device description and intended purpose, including performance characteristicsQ1 (performance characteristics must be documented in the technical file, not just the IFU)
MDR Annex II 6.1(b)Technical documentation must include design and manufacturing information sufficient to understand design stages and V&VQ2, Q3, Q4 (test environment and evidence must be traceable)
MDR Annex II 6.2(f)Technical documentation must include verification and validation results and evidenceQ2 (test environment), Q3 (wrong evidence path), Q4 (evidence mismatch)
EN 62304 §5.5Software integration and integration testing — requires documented test plan including environmentQ2 (test environment), Q3, Q4 (integration test evidence)
EN 62304 §5.6Software system testing — requires documented system test plan and evidenceQ2 (test environment specification)
EN 82304-1 §6.2Health software validation — evidence must include environment specificationQ2 (test environment)

GSPR 15.1 — our position on diagnostic function​

Our GSPR analysis (R-TF-008-001) states that "the device does not have a diagnostic nor a measuring function but it is intended to provide information to support practitioners in decision making for diagnostic purposes." Despite this, GSPR 15.1 is marked as applicable because the device provides information that influences diagnostic decisions.

This position must be proactively clarified in our response. The device is a clinical decision-support tool (Rule 11, MDR Annex VIII), not a diagnostic device. It does not make diagnoses — it provides supplementary information (severity scores, ICD probability distributions, image analysis) that practitioners use alongside their clinical judgement. GSPR 15.1 applies because the accuracy and precision of that supplementary information must be communicated to users so they can weigh it appropriately. Our response should frame the IFU updates as fulfilling GSPR 15.1's requirement to "indicate limits of accuracy" for decision-support information, not as an admission that the device has a diagnostic function.

Response approach for a Major NC​

This is a Major NC. Our responses need to be thorough and demonstrate corrective action. For each question:

  1. State the regulatory requirement being addressed — cite specific GSPR/Annex II clauses
  2. Acknowledge the gap — don't be defensive
  3. Explain what was already in place — demonstrate maturity of the system
  4. Describe corrective action taken — with specific document references
  5. Provide evidence — red-lined documents and supplementary evidence PDF

Cross-cutting: evidence packaging CAPA​

Systemic issue

Questions 3 and 4 both involve incorrect or mismatched evidence in the integration test package submitted to BSI. This suggests a systemic issue with evidence packaging/compilation, not isolated errors. BSI may raise this in Round 2 if we only fix the individual cases.

The systemic issue​

Two out of 76 integration test evidence packages had errors:

  • T355 (Q3): Evidence URI pointed to the wrong S3 path (data entry error in CSV)
  • T377 (Q4): Evidence response.json does not match expected output (structural mismatch or compilation error)

This is a 2.6% error rate in evidence packaging. For a Major NC under Annex II 6.2(f), BSI will expect confidence that the remaining 74 test cases are correct.

Proposed corrective and preventive action​

  1. Audit all 76 evidence URIs against their test case definitions. For each test case, verify:

    • The Evidence URI path contains the correct model name
    • The response.json at that path has the same JSON structure as the expected result in the CSV
    • Numerical values are within the specified acceptance criteria
  2. Add automated validation to the evidence collection pipeline:

    • Schema validation: check that evidence response.json has the same top-level keys as the expected result
    • URI validation: check that the evidence URI path matches the model name in the test case
    • Tolerance check: verify all numerical values are within acceptance criteria
  3. Document the audit results as supplementary evidence (table showing all 76 test cases verified, with pass/fail and any corrections made).

Action items for CAPA​

#ActionOwnerDocument affectedPriority
12Audit all 76 integration test evidence URIs and contentGerardoSupplementary evidence PDFHigh
13Add automated evidence validation to integration pipelineGerardoIntegration test toolingMedium

Summary of action items​

#ActionOwnerDocument affectedPriority
1Add clinical studies reference section to IFUTaig + Jordiifu.pdf (eu-ifu-mdr)High
2Add ICD-11 codes to performanceClaims.ts data modelTaig + JordiperformanceClaims.tsHigh
3Surface ICD-11 codes in IFU performance claimsTaig + JordiClinicalBenefitsList.tsx, ifu.pdfHigh
4Add clinical benefit glossary/legend table to IFUTaigclinical-benefits-and-performance-claims.mdxHigh
5Add consolidated test environment section to Test ReportGerardoR-TF-012-035High
6Verify correct biofilm evidence exists at correct S3 pathGerardo—High
7Fix Evidence URI for T355 in CSVGerardoai-models-integration-tests.csvHigh
8Export correct biofilm/slough evidence from S3GerardoSupplementary evidence PDFHigh
9Add evidence URI validation to integration test pipelineGerardoIntegration test toolingHigh
10BLOCKER: Investigate T377 response.json in S3Gerardo—Critical
11Provide corrected T377 evidenceGerardoSupplementary evidence PDFHigh
12Audit all 76 integration test evidence URIs and contentGerardoSupplementary evidence PDFHigh
13Add automated evidence validation to integration pipelineGerardoIntegration test toolingMedium
14Compile red-lined documentsTaigBSI response packageHigh
15Compile supplementary evidence PDF with bookmarksTaigBSI response packageHigh

Dependencies​

Timeline​

  • Deadline: 2026-03-17 (Round 1 response)
  • Blockers must be resolved by: 2026-03-05 (items 6, 10)
  • Documentation updates: 2026-03-07 (items 1-5, 7)
  • Evidence compilation: 2026-03-10 (items 8, 11, 12)
  • Internal review deadline: 2026-03-14 (items 14, 15)
  • Submission: 2026-03-17
Previous
Technical Review: Round 1
Next
Question
  • Requirements and references
  • Documents reviewed by BSI
  • Regulatory context
    • Cited requirements — detailed breakdown
    • GSPR 15.1 — our position on diagnostic function
    • Response approach for a Major NC
  • Cross-cutting: evidence packaging CAPA
    • The systemic issue
    • Proposed corrective and preventive action
    • Action items for CAPA
  • Summary of action items
    • Dependencies
  • Timeline
All the information contained in this QMS is confidential. The recipient agrees not to transmit or reproduce the information, neither by himself nor by third parties, through whichever means, without obtaining the prior written permission of Legit.Health (AI Labs Group S.L.)