Question

Evidence that integrated model 'Biofilm/Slough Surface' is/has been verified through integration verification is not observed. The test evidence provided in the folder C515_T355 appears to be for wound bed, not biofilm/slough surface. E.g., excerpt from response.json:

[]

Please provide updated evidence.