Question

Test evidence in waf_*.png and test_rail_verification_tests-44bb53ee75c87f0e606ac68be0733274_results.csv for C366/T228 demonstrate parts 1, 2 and 3 of SRS-CCD (traffic inspection, automatic blocking and anomaly detection); however, it is unclear how parts 4 and 5 (alerting and audit logging; fail-safe behaviour) were verified by test T228. In addition, actual results demonstrating that the expected results in test_rail_verification_tests-44bb53ee75c87f0e606ac68be0733274_results.csv for test T228 were achieved (e.g., HTTP 200 and 429 return codes, audit log evidence) could not be found. Please provide the location of this information or provide updated documentation if necessary.

Previous
M2: Software V&V
Next
Research and planning
All the information contained in this QMS is confidential. The recipient agrees not to transmit or reproduce the information, neither by himself nor by third parties, through whichever means, without obtaining the prior written permission of Legit.Health (AI Labs Group S.L.)