Question

Use errors, close calls and difficulties were observed during summative evaluation, HCP use scenario 3. Evidence that they were subjected to root cause analysis and residual risk assessment per Summative Evaluation Protocol, 14.7. Data Analysis and EN 62366-1, 5.9 could not be found.