SRS-027 The product provides a final image validity summary
Identifier
SRS-027
Software system
Backend
Category
- Functional
 - Input and output, interfaces
 
Description
The product shall aggregate the results from all individual image adequacy assessments to produce a final, single boolean isValid flag.
This flag shall be set to true if, and only if, all of the following conditions are met:
- The image quality assessment (per SRS-024) results in an 
acceptableflag oftrue. - The clinical domain validation (per SRS-025) results in an 
isDermatologicalflag oftrue. 
If any of these conditions are not met, the isValid flag shall be set to false, and no further clinical analysis (e.g., PRS-8QJ, PRS-2ZB) shall be performed on the image.
Derived from PRS
PRS-7XK: Assess image adequacy on ingestion